Mipdatabase live, currently comprising 19367 items



Reference Type: Journal
Authors:   Liu,T.  Wei,X.P.  Li,J.P.    
Article title: An electrochemical molecular imprinted sensor based on silicon dioxide films for oxytetracycline detection
Publication date: 2013
Journal : Journal of Analytical Science
Volume: 29
Issue: (1)
Page numbers: (25-29)
Alternative URL: http://en.cnki.com.cn/Article_en/CJFDTotal-FXKX201301007.htm

Abstract: A molecular imprinted membrane for oxytetracycline recognition was prepared by chemical polymerization using 3-aminopropyl trimethoxy siloxane (APTMS) as the functional monomer, tetraethyl orthosilicate (TEOS) as the cross-linking agent and ammonia as the initiator. An electrochemical sensor for oxytetracycline was constructed by coating t he molecular imprinted membrane on the surface of an Au electrode. A method for oxytetracycline using potassium ferricyanide as the ionic probe was established with cyclic voltammetry and differential pulse voltammetry. A linear response was obtained within the oxytetracycline concentration range of 3.0 × 10-7 - 4.5 × 10-6 mol/L with a detection limit of 3.2 × 10-8 mol/L. The recoveries in milk samples were 99.4%-101.1%
Template and target information:   oxytetracycline




Disclaimer: While every effort has been made to provide accurate information, users are recommended to verify data at the publisher's website

Note: The database is currrently being brought up to date and many articles are therefore not included at this time, please bear this in mind when searching for papers with publication dates from 2022 onwards. Current coverage extends from 2005 to date, for older data, please consult the original mipdatabase website