Mipdatabase live, currently comprising 19367 items



Reference Type: Journal
Authors:   Moreira,F.T.C.  Dutra,R.A.F.  Noronha,J.P.C.  Fernandes,J.C.S.  Sales,M.G.F.    
Article title: Novel biosensing device for point-of-care applications with plastic antibodies grown on Au-screen printed electrodes
Publication date: 2013
Journal : Sensors and Actuators B: Chemical
Volume: 182
Page numbers: (733-740)
DOI: 10.1016/j.snb.2013.03.099
Alternative URL: http://www.sciencedirect.com/science/article/pii/S0925400513003882

Abstract: A gold screen printed electrode (Au-SPE) was modified by merging Molecular Imprinting and Self-Assembly Monolayer techniques for fast screening cardiac biomarkers in point-of-care (POC). For this purpose, Myoglobin (Myo) was selected as target analyte and its plastic antibody imprinted over a glutaraldehyde (Glu)/cysteamine (Cys) layer on the gold-surface. The imprinting effect was produced by growing a reticulated polymer of acrylamide (AAM) and N,N'-methylenebisacrylamide (NNMBA) around the Myo template, covalently attached to the biosensing surface. Electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV) studies were carried out in all chemical modification steps to confirm the surface changes in the Au-SPE. The analytical features of the resulting biosensor were studied by different electrochemical techniques, including EIS, square wave voltammetry (SWV) and potentiometry. The limits of detection ranged from 0.13 to 8 μg/mL. Only potentiometry assays showed limits of detection including the cut-off Myo levels. Quantitative information was also produced for Myo concentrations GëÑ 0.2 μg/mL. The linear response of the biosensing device showed an anionic slope of ~70 mV per decade molar concentration up to 0.3 μg/mL. The interference of coexisting species was tested and good selectivity was observed. The biosensor was successfully applied to biological fluids
Template and target information:   myoglobin; Myo
Author-assigned keywords: Surface molecular imprint; self-assembled monolayer; Screen-printed electrodes; Cardiac biomarker; myoglobin; biosensor




Disclaimer: While every effort has been made to provide accurate information, users are recommended to verify data at the publisher's website

Note: The database is currrently being brought up to date and many articles are therefore not included at this time, please bear this in mind when searching for papers with publication dates from 2022 onwards. Current coverage extends from 2005 to date, for older data, please consult the original mipdatabase website