Abstract: The authors describe a microparticle-based system for the detection of the fluoroquinolone antibiotic ciprofloxacin. The method is using the tris(dibenzoylmethane)(1,10-phenanthroline)europium(III) luminophore in polystyrene microparticles along with a molecularly imprinted polymer (MIP) for ciprofloxacin. If ciprofloxacin is captured by the MIP, it quenches the fluorescence of the luminophores. Fluorescence drops linearly in the 0.5-100 μg L-1 ciprofloxacin concentration range, and the detection limit is 92 ng L-1. The method was applied to the analysis of fish samples to assess the analytical performance of the probe. Recoveries ranged from 85.4 to 86.6%, and relative standard deviations between 2.1 and 3.9% (for n = 5)
Template and target information: ciprofloxacin
Author keywords: Fluorometric determination, Polystyrene microparticles, molecularly imprinted polymer, Europium(III)(DBM)3Phen, Luminophore, Ciprofloxacin, Fluorescent probe, fish sample, selective recognition, Quenching