Abstract: A monolithic solid-phase microextraction (SPME) fiber was fabricated based on a molecularly imprinted polymer which could be coupled with gas chromatography for extraction, pre-concentration and determination of methamphetamine (MAMP). Methacrylic acid, ethylene glycol dimethacrylate and MAMP play the roles of functional monomer, cross-linker and template, respectively. The effective factors influencing the polymerization and extraction procedures were investigated and will be detailed here. The fabricated fiber was firm, inexpensive, stable, selective and durable which gives it vital importance in SPME. Selectivity of the fabricated fiber in relation to the template in solution containing MAMP, related and unrelated compounds was also investigated. Under the optimum conditions, the calibration plot was linear in the range of 50 - 3,500 ngmL-1 (r 2 =0.997). The high extraction efficiency was obtained for MAMP giving a detection limit of 14ngmL-1. The fabricated fiber was successfully applied to SPME of MAMP from human saliva samples followed by gas chromatography-flame ionization detector analysis
Template and target information: methamphetamine, MAMP
Author keywords: gas chromatography, Methamphetamine, molecularly imprinted polymer, Saliva, solid-phase microextraction